GUI reference
R&S
®
ZNA
556User Manual 1178.6462.02 ─ 20
Basic mode
In basic mode, the validation runs automatically, with autodetected ports and the set-
tings specified in Advanced mode (default settings after a system [Preset]).
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Manual port assignment is not supported.
Validate
Runs the validation for the current port assignment. The validation proceeds through
the connected ports p ∈ {1, ..., 4} in ascending order, measuring S
pp
for the Open,
Short and Match standard.
●
"PASS": all measured reflection parameters are within the specified Validation Lim-
its, compared to the selected reference characterization.
●
"FAIL" otherwise.
[In this case, the validation stops at the first failed reflection measurement.]
Remote command:
CALCulate:CALValidate:RUN
Advanced Mode
Switches to Advanced mode.
Advanced mode
In advanced mode you can configure the validation settings (Characterization, Valida-
tion Limits) and run or rerun particular measurements (Measurement, "Calibration
Standard" on page 557 ).
Cal softtool