Measurements and Result Displays
R&S
®
ESW
134User Manual 1177.6298.02 ─ 08
The list contains all frequencies that have been measured during the final measure-
ment.
Figure 9-15: Overview of the "Final Results" table
The functionality and information provided by the final results dialog box is similar to
that provided by the peak list.
Functions to manage final results described elsewhere:
"Sorting the peak list" on page 128
"Exporting a peak list" on page 129
"Displaying peaks as symbols" on page 129
Alternate way to modify the final results
Like the peak list, you can also modify the final results directly via the "Final Meas"
result display.
More information.
Select the "SmartGrid" icon (
) and start the "Final Meas" result display.
The remote commands required to manage final results are described in Chap-
ter 15.5.9, "Final Measurement (and Trace) Configuration", on page 472 and Chap-
ter 15.5.10, "Final Results", on page 475.
Reading the peak list
The table at the top of the dialog contains information when you measure with limit
lines. Each cell of the table represents a trace and shows the limit lines that are
assigned to that trace.
The main table shows the following information for each frequency that has been mea-
sured. The size of the table depends on the number of peaks that have been found.
Rows with no highlighting
Limit check for the corresponding frequency has passed.
Rows in red highlighting
Test Automation