General Instrument Setup
R&S
®
ESW
303User Manual 1177.6298.02 ─ 08
Self-alignment results in secure user mode
Be sure to store self-alignment results before SecureUser Mode is enabled; see Chap-
ter 4.7, "Protecting Data Using the Secure User Mode", on page 40.
In secure user mode, the results are not stored permanently. Thus, if the currently
stored self-alignment results are not suitable, you must perform a self-alignment each
time you switch on the R&S ESW.
Start Self Alignment.................................................................................................... 303
Abort Self Alignment................................................................................................... 303
Await Warm-Up Operation before Self Alignment....................................................... 304
Shut down Device after Self Alignment.......................................................................304
Starting Touch Screen Alignment................................................................................304
Alignment Results:...................................................................................................... 304
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see Chapter 13.1.1, "Basics on Alignment", on page 300.
Note:
A running Sequencer operation is aborted when you start a self-alignment.
Remote command:
*CAL? on page 431, see also CALibration[:ALL] on page 625
Abort Self Alignment
As long as the self-alignment data is being collected, the procedure can be canceled
using the "Abort Self Alignment" button.
Note: If you start a self-alignment remotely, then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed. In this case, you cannot abort a self-alignment man-
ually.
Alignment