Remote Commands in the Receiver Application
R&S
®
ESW
480User Manual 1177.6298.02 ─ 08
Suffix:
<sr>
.
irrelevant
Parameters:
<Phase> You can select several phases for the scan:
<Phase>,[<Phase>,<Phase>,<Phase>]
L1
L2
Available for networks with four phases (R&S ESH2Z5,
R&S ENV4200 and R&S ENV432)
L3
Available for networks with four phases (R&S ESH2Z5,
R&S ENV4200 and R&S ENV432)
N
*RST:
L1
Example: //Select phases to be measured in the scan
SCAN:LISN:PHAS L1,N
Manual operation: See "Phase" on page 136
INPut<ip>:LISN:FILTer:HPASs[:STATe] <State>
This command turns the 150 kHz highpass filter of the ENV216 network on and off.
Prerequisites for this command
Select ENV216 network (INPut<ip>:LISN[:TYPE]).
Suffix:
<ip>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on high pass filter
INP:LISN:TYPE ENV216
INP:LISN:FILT:HPAS ON
Manual operation: See "High-Pass Filter 150 kHz" on page 137
INPut<ip>:LISN:PHASe <Phase>
This command selects one LISN phase to be measured.
The command is available in all applications (spectrum, receiver, I/Q analyzer and ana-
log demodulator).
Suffix:
<ip>
.
irrelevant
Parameters:
<Phase> L1
Measurements and Result Displays