Remote Commands in the Receiver Application
R&S
®
ESW
521User Manual 1177.6298.02 ─ 08
To add a new window, use the LAYout:ADD[:WINDow]? command.
Setting parameters:
<WindowName> String containing the name of the existing window.
By default, the name of a window is the same as its index. To
determine the name and index of all active windows in the active
channel, use the LAYout:CATalog[:WINDow]? query.
<WindowType> Type of result display you want to use in the existing window.
See LAYout:ADD[:WINDow]? on page 518 for a list of availa-
ble window types.
Example:
LAY:REPL:WIND '1',MTAB
Replaces the result display in window 1 with a marker table.
Usage: Setting only
LAYout:SPLitter <Index1>, <Index2>, <Position>
This command changes the position of a splitter and thus controls the size of the win-
dows on each side of the splitter.
Note that windows must have a certain minimum size. If the position you define con-
flicts with the minimum size of any of the affected windows, the command will not work,
but does not return an error.
Figure 15-1: SmartGrid coordinates for remote control of the splitters
Setting parameters:
<Index1> The index of one window the splitter controls.
<Index2> The index of a window on the other side of the splitter.
Analysis