Remote Commands in the Receiver Application
R&S
®
ESW
637User Manual 1177.6298.02 ─ 08
Example:
SENSe1:CORRection:TRANsducer:UNIT 'DB'
// Frequency Span 0 Hz to 4 Ghz
SENSe1:CORRection:TRANsducer:DATA 0,8,2GHz,5,4GHz,3
Creates the transducer points:
Frequency Level
0 Hz 8 dB
2 GHz 5 dB
4 GHz 3 dB
[SENSe:]CORRection:TRANsducer:DELete
This command deletes the currently selected transducer factor.
Before you can use the command, you have to select a transducer.
Example:
CORR:TRAN:DEL
[SENSe:]CORRection:TRANsducer:SCALing <ScalingType>
This command selects the frequency scaling of the transducer factor.
Parameters:
<ScalingType> LINear | LOGarithmic
*RST: LINear
[SENSe:]CORRection:TRANsducer:SELect <Name>
This command selects a transducer factor.
Parameters:
<Name> String containing the name of the transducer factor.
If the name does not exist yet, the R&S ESW creates a trans-
ducer factor by that name.
Example:
CORR:TRAN:SEL 'FACTOR1'
[SENSe:]CORRection:TRANsducer:INPut<rf>[:STATe] <State>
[SENSe:]CORRection:TRANsducer[:STATe] <State>
This command turns the selected transducer factor on or off.
Before you can use the command, you have to select a transducer.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0