Remote Commands in the Receiver Application
R&S
®
ESW
659User Manual 1177.6298.02 ─ 08
Suffix:
<j>
.
Selects the index of a custom synchronization mechanism.
You can query the index with INSTrument:COUPle:USER<j>:
NUMBers:LIST?.
Parameters:
<Scope> ALL
Optional SCPI parameter.
Deletes all custom couplings.
You do not have to use the suffix <j> in that case.
Example:
INST:COUP:USER3:REM
Removes the custom coupling with the index number 3.
Usage: Event
INSTrument:COUPle:USER<j>:STATe <State>
This command turns a custom coupling mechanism on and off.
Suffix:
<j>
.
Selects the index of a custom synchronization mechanism.
You can query the index with INSTrument:COUPle:USER<j>:
NUMBers:LIST?.
Parameters:
<State> ON | OFF
Example:
INST:COUP:USER2:STAT ON
Turns on the coupling with the index number 2.
INSTrument:COUPle:USER<j>:WINDow:LIST? [<ChannelName>, <Parameter>]
This command queries the measurement windows that can be synchronized with
another channel (or measurement window).
Note that synchronizing with a specific measurement window is only possible in Analog
Demodulation application.
Suffix:
<j>
.
irrelevant
Query parameters:
<ChannelName> Optional SCPI parameter.
String containing the name of a measurement channel.
<Parameter> Optional SCPI parameter.
String containing the name of a parameter that you can syn-
chronize.
Return values:
<Synchronizable
Window>
Case 1 (command without SCPI parameters):
Always returns the string 'All Windows'.
Case 2 (command with SCPI parameters):
General Instrument Setup