5. Application Operation
5-114
CBM-20A/20Alite
1. The maximum pressure limit for the entire system (system P.Max) can be set. This setting is provided
to avoid applying a pressure greater than the allowable withstand pressure to the connected
autosampler. This setting is available only when autosampler is connected. When the system P.Max
setting is enabled, an error will occur in the event that the pressure has risen above the system P.Max
value.
2. The system P.Max value varies depending on the connected autosampler. The table below gives the
system P.Max values for respective models. If no autosampler is connected, the system P.Max value
defaults to 0.
Miscellaneous
Shut down Multiple
Terminal Box on error
Disable
Enable
Disable
Disable:Power stays ON when an error
occurs.
Enable:Power turns OFF when an error
occurs.
Control Mode
Normal
Fast LC
Normal
Normal: Control HPLC units as the normal
mode.
Fast LC: Select to analyze with the Fast LC
units.
Model name System P.Max value (MPa)
SIL-30ACMP
SIL-30AC
130.0
SIL-20AXR
SIL-20ACXR
66.0
SIL-20AHT UFLC
SIL-20ACHT UFLC
SIL-20AHT
SIL-20ACHT
35.0
SIL-20A
SIL-20AC
20.0
SIL-10ADvp
SIL-10A
SIL-10AF
SIL-10AP
34.3
SIL-10Ai 19.6
Not connected 0
Parameter
Setting
range
Default
setting
Remark

Table of Contents

Other manuals for Shimadzu CBM-20A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu CBM-20A and is the answer not in the manual?

Shimadzu CBM-20A Specifications

General IconGeneral
BrandShimadzu
ModelCBM-20A
CategoryController
LanguageEnglish

Related product manuals