9.3 Initializing Components
9-31
CBM-20A/20Alite
9
4
Enter a number from 1 to 4 to change the corresponding setting.
Refer to "Parameter Settings" P. 9-31 for more details.
Select 0 to save the current settings and start the system.
Parameter Settings
Parameter Description
1. IP address
Set the IP address and the subnet mask.
Set [0.0.0.0] for the IP address when obtaining an IP address
automatically from a DHCP server.
2. Subnet mask
3. Default gateway
Sets the default gateway.
If there is no default gateway, input [-].
4. Ether speed
Sets the Ethernet speed/mode for the CBM-20A.
Any value in the range 0 to 4 can be set.
0 : Auto
1 : 10M/Half
2 : 10M/Full
3 : 100M/Half
4 : 100M/Full
Normally, the setting "0: Auto" (automatic switching) is used. If, however,
it is not possible to link to the hub of the communications destination, the
Ethernet speed/mode can be fixed by setting a value in the range 1 to 4.
5. Boot parameter
Do not use these settings. (For service purpose only)