Page EN-6 XRS-5 Series Transducers – OPERATION MANUAL 7ML19985QT81
mmmmm
English
Interconnection
Recommended installation rules:
Do not route cable openly; instead, run cable separately in a grounded metal conduit to
protect it from ambient electrical noise.
Seal all thread connections to prevent the ingress of moisture.
Do not run cable near high voltage or current runs, contactors or SCR control drives.
In all of the following examples the terminal blocks on the transceiver are described in the
controller manual.
Blind Flange
Face Flange
Note: Tighten the flange bolts evenly in order to ensure a good seal between the mating
flanges
.
Caution: Over-tightening can cause performance degradation.
Note: Installation shall only be performed by qualified personnel and in accordance with
local governing regulations.
nipple welded to blind flange
coupling
coupling
bolt
gasket
customer flanged,
flat face only
nut
factory flanged
transducer