SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
257
Reminder Activation (3.3.6.4.)
Note
To modify this parameter via SIMATIC PDM it must be accessed via the pull
-down menu
Device
Maintenance.
Allows you to enable a maintenance reminder.
Values
*
TIMER OFF
ON NO LIMITS - no reminders checked
ON REMINDER 1 (MAINTENANCE REQUIRED) checked
ON REMINDER 1 AND 2 checked
ON REMINDER 2 (MAINTENANCE DEMANDED) checked
1. First set the values in Reminder 1 before Calibration (Required) (3.3.6.5.)
(Page 257)/Reminder 2 before Calibration (Demanded) (3.3.6.6.) (Page 257).
2. Select the desired Reminder Activation option.
Reminder 1 before Calibration (Required) (3.3.6.5.)
If Time Next Calibration (3.3.6.3.) (Page 256) is equal to or less than this value, the device
generates a Maintenance Required reminder.
Values
Range: 0.000 to 20.000 years
Default: 0.164 years (8 weeks)
1. Modify values as required.
2. Set Reminder Activation (3.3.6.4.) (Page 257) to the desired option.
Reminder 2 before Calibration (Demanded) (3.3.6.6.)
If Time Next Calibration (3.3.6.3.) (Page 256) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
Values
Range: 0.000 to 20.000 years
Default: 0.019 years (1 week)
1. Modify values as required.
2. Set Reminder Activation (3.3.6.4.) (Page 257) to the desired option.
Maintenance Status (3.3.6.7.)
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu View Device Status, click on the Maintenance tab and
check the Calibration Schedule Status window.