EasyManua.ls Logo

Sonardyne 8270 - Section 8 - Fault Diagnosis; Introduction; Test Equipment; Quick Voltage Checks

Sonardyne 8270
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WSM6 Manual
Document Ref: UM-8270-000
Issue: A Rev 0
Section 8
89 October 2011
SECTION 8 – FAULT DIAGNOSIS
8. Introduction
The following section assists in diagnosis of a WSM6 when fully disassembled. For instructions on
disassembly refer to Section 6.5Dismantling.
8.1 Test Equipment
To assist in determining a fault with the WSM6 it is recommended to have the following list of
equipment:
Sonardyne ANT Type 7343 – firmware version 2.25 or greater, with ANT coupling loop
7343-042B or ANT test transducer Type 7935-000-B.
Bench Power Supply Unit – which can be set to any output voltage over a range up to
at least 25V dc and with a current limit facility which can be set to 100mA.
Digital Multimeter – capable of reading at least 25V and 100mA dc.
Dual Channel Oscilloscope – minimum 10MHz bandwidth, 10 mV/division minimum
sensitivity, with a x10 probe plus a x100 probe capable of use up to 4kV. A guarded
probe tip for connection to integrated circuit legs is strongly recommended.
Frequency Counter – capable of measuring up to 200kHz minimum, accuracy 0.1%.
Test Cable 1 – a Type MCIL5F connector with 300cm tail.
Test Cable 2 – PSU (18V) to the battery connector (4-pin Molex, pin 2 = +, pin 4= -)
8.2 Quick Voltage Checks
1. With the unit switched ON, measure the battery voltage using a Digital Multimeter, across the
terminal of the large capacitor – C16 on the Tx board.
2. Measure the voltage of the CPU logic rail by moving the positive probe to pin 3 of JP3, which
should read 5V ± 0.1V. JP3.3 is the unsoldered pad of JP3, closest to the central fixing screw
on the upper board, see Figure 8-3.
3.
Measure the 5V rail on the Tx card which is generated by a separate regulator, on the via
hole closest to TP3 on the lower board, refer to Figure 8-2.
8.3 Diagnostic Set-up
1. Check the ON/OFF switch is in the OFF position. Make sure the Power Supply Unit (PSU) is
OFF. Refer to Figure 8-1.
2.
Disconnect the battery pack from the Tx card by disconnecting PL1, refer to Figure 8-2.
3.
Connect Test Cable 2 to PL1 on the Tx card, refer to Figure 8-2.
4.
Connect Test Cable 2 Molex pin 4 to the PSU –ve terminal.
5. Connect an ammeter (100mA dc range) between the PSU +ve terminal, and Molex pin 2.

Table of Contents

Related product manuals