7-15
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 33
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 33 =
BOOT B: HIGH PRESSURE
Possible causes:
1. An endoscope of a diameter larger than template is
processed with insertion tube sleeve(s). Do not use
insertion tube sleeve for these endoscopes (refer to
S
ECTION 4.3).
2. Relief valve is not working properly. Call STERIS.*
3. Relief valve pressure incorrectly adjusted. Call STERIS.*
4. If 60 Hz processor, ensure there is a spacer between
valve cap and valve stem.
5. Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 34
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 34 =
BOOT A: OVER PRESSURE
This message warns operator that high pressure in control
handle boot "A" could damage endoscope.
IMPORTANT: If high level disinfection solution remains in
sump, call STERIS.*
Possible causes:
1. An endoscope of a diameter larger than template is
processed with insertion tube sleeve(s). Do not use
insertion tube sleeve for these endoscopes (refer to
S
ECTION 4.3).
2. Relief valve is not working properly. Call STERIS.*
3. Relief valve pressure incorrectly adjusted. Call STERIS.*
4. For a 60 Hz processor, ensure there is a spacer between
valve cap and valve stem.
5. Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION