4.13. Synchronization
• Update I™ / synchronize data upon connection:
Set/Check this feature to enable automatic synchronization and updat-
ing for the I™.
• Set time:
Set/Check this feature to set the clock-time on the unit to the PC
clock-time upon synchronization. Make sure that the date and time
are correct on the PC prior to synchronization.
4.13 Synchronization
The Synchronization process includes the following information:
• I™ device settings
• recorded spectra
• event log
• history log
• recorded voice and image files
The device parameters must be loaded before synchronizing between the unit
and the PC. Any connected device requiring a parameter download is shown
with a “stop hand” symbol in front of its label, while one that is ready for
usage or synchronization is shown with a circular, green arrow. The simulta-
neous download of device parameters for multiple units within a group is not
possible: this process activates for one unit at a time.
Warning Upon synchronization, all data from the unit will be transferred
to the PC and deleted in order to free memory! Data cannot be “transferred”
back to the device. s
Warning Do not press any of the buttons on the I™ during
unit synchronization uploads and downloads to and from the PC. Doing so
can lock up the unit. s
Thermo Fisher Scientific Interceptor™/en/5.1(4956)/Feb2010 87