4.12. Device Configuration
4.12.1 Gamma Settings
Warn Level
Enter a value between 0.25 to 1,000.0 µSv/h or 25.0 to 100,000.0
µRem/h for the Warn Level. This defines the absolute threshold for
audible and visual warning events. The factory default value is 400
µRem/h.
Alarm Level
Enter a value between 0.25 to 1,000.0 µSv/h or 25.0 to 100,000.0
µRem/h for the Alarm Level. This value defines the absolute threshold
for audible and visual alarming events. The factory default value is
4,000 µRem/h.
Sigma Alarm Level
Enter a value between 3.0 and 9.9 for the Sigma Alarm Level. This
value defines the floating threshold over background for audible
and visual alarm events. The factory default value is 6 (i.e. less than
0.0004% chance that an alarm activates under a non-alarm level
condition). The Sigma Alarm can also be disabled by unchecking the
checkbox beside it.
Speaker
Turns the speaker for gamma events on or o.
Vibrator
Turns the vibrator for gamma events on or o.
Units
Select the units of measurement to be displayed on the device.
Callback Number / Identification
A phone number which is displayed with the Isotope descriptions in
ID Mode.
Wake on Sig
When checked, the display will awaken upon sigma alarm, gamma
warning, or gamma alarm. When unchecked, the display will only
awaken on gamma warnings or gamma alarms.