Appendix C
Electrical Test Results
Thermo Fisher Scientific Corporation declares under their sole responsibility
that the product I™ is in conformity with the following stan-
dards or other normative documents:
Conducted Emissions (EN 55011 and FCC Part 15) Passed
Radiated Emissions (EN 55011 and FCC Part 15) Passed
Harmonics (EN 61000-3-2) N/R *
Flicker (EN 61000-3-3) N/R *
Electrostatic Discharge (IEC 61000-4-2) Passed
Radiated Immunity (EN 61000-4-3) Passed
Fast Transient Burst (EN 61000-4-4) Passed
Surges (IEC 61000-4-5) Passed
Conducted Immunity (IEC 61000-4-6) Passed
Power Frequency Magnetic Field (IEC 61000-4-8) Passed
Voltage Dips and Interrupts (IEC 61000-4-11) Passed
N/R * Power Consumption less than 50 Watts
Test Report No. G610211
This report concerns EMC (ref. IEC 61000-6-2 (1999)/EN 61000-
6-4 (2001) and FCC Part 15) Testing of In-
dustrial Equipment
The test report was
prepared by and test
performed at
GLOBAL TESTING LABORATORIES,
LLC
3029 East Governor John Sevier Highway
Knoxville, Tennessee 37914-6424
October–November 2006
Thermo Fisher Scientific Interceptor™/en/5.1(4956)/Feb2010 127