G. Frequently Asked Questions (FAQ)
the low-pressure neutron detector would be identified as either a GNid
+L +C +BT or a GN +L +C +BT.
Q. Will Reach-Back work with all wireless telephones?
A. No... In order to be compatible with Reach-Back, the wireless phone
must have Bluetooth® technology, an internal modem, and dial-up
networking (DUN) in order to access a Reach-Back server.
Q. Should the device be returned once periodically for calibration?
A. ICx Radiation, recommends returning the InterceptorTM annually for
optimization and once every 2 years to have the battery replaced.
Q. Can spectra be downloaded from the Interceptor TM using the
winTMCA32 software?
A. No... this software package is used for further analysis of spectra.
Q. What do the functional test results indicate?
A. See chart below:
Functional Test Description
High Voltage Test Verifies the detectors voltage is
within tolerance
Spectra Test No test
Device Communication Test Checks the communications
during the test command
Battery Communication Test Checks the battery monitor
communications which is
where the Battery Gas Gauge
percentage is derived
Battery Voltage Test Checks the battery voltage to
make sure it is between the
Max (4.6V) and Min (2.9V)
limits
to be continued. . .
Thermo Fisher Scientific Interceptor™/en/5.1(4956)/Feb2010 139