6 Create Methods, Sample Sets, and Sample Templates
152 July 29, 2013, 715003794 Rev. B
The following table describes the fields and screen keys in the Mobile Phase
screen.
Mobile Phase screen parameters:
Parameter Function Value range
Initial Flow Specifies the initial flow rate of the
method. For isocratic operation, this
is the flow rate for the entire
separation.
0.000 and 0.010 to
10.000 in
0.001 mL/min
increments
Initial
Composition
Specifies the initial composition of
the eluents. The sum of the four
fields must equal 100%. (You enter
solvent names using the Labels
screen key.)
0 to 100.0 in 0.1%
increments
Alarms: Min Specifies the system pressure (set in
psia, bar, or kPa) below which the
alarm condition (selected in the
adjacent box) is executed. To enable
access to the Pressure fields, set the
alarm parameter to anything other
than “Disable.”
0 to 4500 in 1 psi
increments, 0 to
310 bar in 1 bar
increments, or 0 to
31025 kPa in 1 kPa
increments
See “Alarms” on
page 154.
Alarms: Max Specifies the system pressure (set in
psia or bar) above which the alarm
condition (selected in the adjacent
box) is executed. Use to detect
problems with method conditions and
to protect your column from
overpressure.
0 to 5000 in 1 psi
increments, 0 to
344 bar in 1 bar
increments, or 0 to
34473 kPa in 1 kPa
increments
See “Alarms” on
page 154.
Alarms: Bubble
Detect
Specifies the response that occurs
when the solvent management
system detects a bubble in the flow
path.
See “Alarms” on
page 154.