6 Create Methods, Sample Sets, and Sample Templates
154 July 29, 2013, 715003794 Rev. B
Alarms
The separations module maintains a log of all enabled errors that occur
during operation. Each error can produce a variety of responses, as shown in
the following table.
Labels
(screen key)
Displays the Solvents dialog box,
which allows you to add, remove, or
change the names of solvents used in
methods.
Use the Add, Remove, and Change
screen keys to edit the list of
solvents.
A, B, C, and D are not valid
user-entered solvent names.
Strk Vol
Seal Wsh
(screen key)
The Preferred Stroke Volume field
sets the volume of solvent delivered
with each piston stroke.
You can override the default stroke
volumes, but do not exceed the flow
rate limits displayed on the screen
for each of the stroke volume
settings. See “Preferred plunger
stroke volume” on page 29.
The Plunger Seal Wash period sets
the time interval between successive
plunger seal-wash pump cycles.
130 µL - default
100 µL
50 µL
25 µL
Off, 0.50 to 10.00 in
0.01 min
increments
Alarm responses:
Alarm response Function
Disable All alarm response reporting is disabled.
Log Quietly The error is entered into the error log without alerting
you.
Mobile Phase screen parameters: (Continued)
Parameter Function Value range