EasyManua.ls Logo

Waters Alliance e2695 - Page 227

Waters Alliance e2695
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Perform main diagnostic tests
July 29, 2013, 715003794 Rev. B 227
Confirm the seal data values—top, bottom, and width—fall within the
acceptable ranges shown in the table, below. A seal geometry error can appear
if these conditions apply:
The adjust seals parameter values are not in the acceptable range.
Air bubbles occur in the sample management system. (If so, purge the
sample management system and try again.)
Solvent is not properly degassed (even though air bubbles are not
visible).
Adjust seals failure
To correct for a failing result on the adjust seals test:
1. Confirm that the flow rate for the adjust seals test was set to 1.0 to 1.5
mL/min, the 100% methanol was properly degassed and filtered, and
that degassing was enabled.
2. If all the conditions in step 1 are satisfied, proceed to step 3.
If any conditions in step 1 are not satisfied, correct it (or them), and then
repeat the test.
Requirement: If the test fails again, proceed to step 3.
3. Replace the restrictor in the red outlet line with a pin plug.
4. Ramp system pressure up to 31026.4 kPa (310.3 bar, 4500 psi); hold this
pressure for 2 minutes.
5. Repeat the adjust seals test.
6. If the test fails again:
a. Stop solvent flow.
b. Monitor the displayed pressure value (which must be stable and
approaching zero).
Acceptable ranges for the adjust seals diagnostic test:
Parameter Acceptable range Typical range
Top 30 to 80 30 to 60
Bottom 80 to 200 120 to 150
Width (Bottom–Top) 50 to 120 60 to 90

Table of Contents

Related product manuals