Connect tubing and attachments
July 29, 2013, 715003794 Rev. B 51
• Replace the feet on the detector and then install the bottle tray and
detector base.
Important: You should replace the feet on the bottom of the detector so
that they correctly fit into the detector base. Failure to do so may result
in the feet becoming wedged or stuck in the detector base.
• Install the detector drip tray tubing.
Required materials
• Diagonal cutters
• Screwdriver, T-20 TORX
To replace the feet and install the bottle tray/detector base:
1. Disconnect all cabling and other connections from the detector.
2. Tip the detector so that it rests on its side.
3. Using a T-20 TORX screwdriver, remove the securing screws and rubber
feet from the chassis.
4. Secure the new plastic feet (410004802) to the chassis using the
provided T-20 screws (WAT034346), and set the detector aside
temporarily.
5. Place the bottle tray on top of the separations module as shown in figure
“B-series solvent bottle tray configuration (only single detector shown):”
on page 25.
6. Position the detector base to the right of the bottle tray, ensuring that
the base is seated correctly on the two locating guides on the right side of
the bottle tray assembly. These locating guides fit into slots on the
underside of the detector base.
7. Ensure that the drip tray is seated in the correct position at the front of
the detector base.
To install the detector drip tray tubing:
1. Prepare and fit the drip tray tubing for your specific model of detector.
The drain fitting for all detectors is located on the underside of the