2 Install the e2695 Separations Module
56 July 29, 2013, 715003794 Rev. B
3. Using the hardware provided with the support bracket, secure the
bracket to the top of the rear panel on the separations module:
• On the left side of the separations module, fasten the two M4 star
screws.
• On the right side of the separations module, fasten the two flat
washers with M5 nuts.
4. Place the detector on the detector support bracket. Make sure that the
rear feet on the bottom of the detector are seated in the large receptacles
in the tabs on either end of the detector support bracket.
To route the drip tray or detector drain tubing into the separations module:
1. For detectors with a drip tray installed (2424, 2489, and 2998 only),
connect one end of a length of convoluted tubing supplied in the kit to
the elbow fitting on the right side of the drip tray.
For detectors without a drip tray, connect one end of the convoluted
tubing to the drain fitting located on the front left-hand side of the
underside of the detector.
2. Route and secure the tubing in the clips in the syringe compartment so
that the end of the tubing is positioned just above the small drip tray
under the syringe, as shown in the figures below.
Note: Pull the tubing tight in the clips, to ensure there are no upward
bends that can cause traps when solvent is draining.