WM 68401 04/2021
MEDUCORE Standard
2
EN 111
5 Function check
The status report appears (example: Function check passed):
20. Print the function check report by pressing the Print function
button (only with Printing option) (see “6.16.5 Printing a
function check report”, page 201).
21. Upload the function check to WEINMANN Connect with the
Upload function button (see “5.3 Uploading function check to
WEINMANN Connect”, page 112).
22. Proceed with the device according to the following table:
Risk of injury due to device not ready for use!
If you operate the device after a failed function check, the patient
may be injured.
Only operate the device after it passes a function check.
Display Meaning Action
Device ready for
use
Function check passed
Use device without
restriction.