WM 68401 04/2021
MEDUCORE Standard
2
EN 135
6 Operation
Shock not possible
The device carries out a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
If, based on the cardiac rhythm analysis, the device determines that
a shock is not required, the device outputs the message:
1. Carry out CPR (see “ Carrying out CPR”, page 135).
Result The patient does not have a cardiac rhythm which can be
defibrillated. The device guides you through CPR by means of voice
prompts, AED instructions, and the metronome. It warns you again
to stand clear of the patient after the preset time has elapsed
(120 s when device supplied), to allow it to carry out another
cardiac rhythm analysis.
Carrying out CPR
This section describes CPR in AED mode. When supplied, the
device carries out CPR with the following parameters; these,
however, can be adapted by the operator:
Voice prompt AED instruction
Stand clear of the patient Stand clear of the patient
Cardiac rhythm is being analyzed Analysis
Voice prompt AED instruction
Shock not recommended Shock not recommended
Setting
Patient group
Adult Child
CPR phase
Duration 120 s 120 s
Pause for ventilation 5 s 5 s