Page #140 background imageLoading...
Page #140 background image
140 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
9. Deliver a shock with the shock button .
When doing so, please note:
If the shock button is not pressed, the shock capacitor
discharges automatically after 30 seconds.
Only with the Print option: If the printer is connected to the
device and the Defibrillation report menu item is
activated (see “13.9 Printer settings (only with Printing
option)”, page 332), the printer automatically prints a
defibrillation report after shock delivery (see “6.16.4
Printing a defibrillation report”, page 201).
Result The patient has been given an electric shock.
6.7.2 Carrying out manual defibrillation using
paddles
This function is only available if manual mode has been enabled
and activated by the operator: Operator menu |
System settings | Enable options | Manual defibrillation (see
“13.12 System settings”, page 342).
If you are the operator of the device and have access to the
operator menu, you can disable manual mode: Operator menu |
System settings | Disable functions | Manual mode (see
“13.12 System settings”, page 342).
Requirement A charged battery is inserted in the device.
The device is switched on (see “6.1 Switching on the device”,
page 118).
A patient group is selected (see “6.4 Selecting patient group”,
page 120).
Failure of therapy due to defibrillation electrodes accidentally
coming loose during shock delivery!
Defibrillation electrodes accidentally coming loose during shock
delivery may lead to damage to the device and thus to the failure
of treatment. This may injure the patient.
Ensure that the defibrillation electrodes are always connected
to the device during shock delivery.

Table of Contents