144 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
13. Press on paddles with a contact pressure of about 8 kg to keep
resistance to the patient low.
14. Press both buttons (SHOCK) on the paddles
simultaneously.
When doing so, please note: Only with Printing option: If the
printer is connected to the device and the Defibrillation
report menu item is activated (see “13.9 Printer settings (only
with Printing option)”, page 332), the printer prints a
defibrillation report (see “6.16.4 Printing a defibrillation
report”, page 201).
Result The patient has been given an electric shock.
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
Stand clear of the patient.
Keep patient away from liquids (e.g. blood, gel or saline
solution).
Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
Keep your distance from liquids in contact with the patient.
Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals