156 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
19. Press and hold both (SHOCK) buttons on the paddles
simultaneously.
When doing so, please note:
The device synchronizes shock delivery with the next R
wave.
Only with Printing option: If the printer is connected to the
device and the Defibrillation report menu item is
activated (see “13.9 Printer settings (only with Printing
option)”, page 332), the printer prints a defibrillation
report (see “6.16.4 Printing a defibrillation report”,
page 201).
After shock delivery, cardioversion (SYNC marking)
remains activated in manual mode. You can set in the
operator menu whether further cardioversion or
defibrillation is to follow cardioversion (see “13.5 Manual
mode settings (only with Manual defibrillation option)”,
page 321).
Result The patient has undergone cardioversion.