6 Operation
1. Connect the SpO
2
connector of the pulse oximetry sensor
connecting cable to the SpO
2
connection on the device.
2. Select the appropriate pulse oximetry sensor for the patient
group:
Only use undamaged pulse oximetry sensors.
Only use the pulse oximetry sensors and pulse oximetry sensor
connecting cables quoted in the scope of supply and in the
accessories.
Pulse oximetry sensor Patient group Application site
MCS2-SoftTip pulse oximetry
sensor, size S
Ø 7.5 mm - 12.5 mm finger diameter
Finger/big toe
MCS2-SoftTip pulse oximetry
sensor, size M
Ø 10 mm - 19 mm finger diameter
MCS2-SoftTip pulse oximetry
sensor, size L
Ø 12.5 mm - 25.5 mm finger diameter
MCS2-Wrap pulse oximetry
sensor
> 10 kg body weight Finger/hand
MCS2-Earclip pulse oximetry
sensor
> 30 kg body weight Ear
MCS2-Wrap pulse oximetry
sensor, adult (Adult), disposable
> 30 kg body weight
Finger/big toe
MCS2-Wrap pulse oximetry
sensor, child (Pediatric),
disposable
10 kg - 50 kg body weight
MCS2-Wrap pulse oximetry
sensor, infant (Infant), disposable