162 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
6. Check whether the oxygen saturation values displayed on the
device are plausible.
Result A pulse oximetry sensor is connected.
6.9.2 Carrying out pulse oximetry monitoring
Requirement The device is switched on (see “6.1 Switching on the device”,
page 118).
A patient group is selected (see “6.4 Selecting patient group”,
page 120).
A pulse oximetry sensor is connected (see “6.9.1 Preparing
pulse oximetry monitoring”, page 157).
1. If the patient group is incorrect: Select another patient group
(see “6.4 Selecting patient group”, page 120).
2. If necessary: Select monitor mode using the Monitor function
button.
3. If necessary: Switch between parameter view and curve view
with the view button .