6 Operation
When doing so, please note:
• The site must have a good blood supply.
• When attaching to the finger, use the ring finger or middle
finger on the non-dominant hand.
• The pulse oximetry sensor must not be attached too tightly.
• The pulse oximetry sensor must be checked every 4 hours
and repositioned if necessary.
• The pulse oximetry sensor must be repositioned if there are
any skin changes.
MCS2-Earclip
pulse oximetry
sensor
MCS2-Wrap
pulse oximetry
sensor,
disposable
Special feature:
The transmitter and
receiver of the
disposable pulse
oximetry sensor
must be aligned to
face each other on
one axis.
Pulse oximetry
sensor
Point of attachment