168 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
8. If ECG electrodes are used at the same time as defibrillation
electrodes: Do not allow ECG electrodes and defibrillation
electrodes to overlap.
9. If necessary: Stroke out any air trapped under the ECG
electrodes.
10. Clip the ECG cable to the individual ECG electrodes.
11. Check whether the ECG curves for ECG recording displayed on
the device are plausible.
Result The 6-lead ECG cable and the ECG electrodes are connected. 6-
lead ECG monitoring is prepared.
6.10.2 Carrying out 6-lead ECG monitoring
Requirement • The device is switched on (see “6.1 Switching on the device”,
page 118).
• A patient group is selected (see “6.4 Selecting patient group”,
page 120).
• 6-lead ECG monitoring is prepared (see “6.10.1 Preparing
6-lead ECG monitoring”, page 164).
1. If the patient group is incorrect: Select another patient group
(see “6.4 Selecting patient group”, page 120).
2. If necessary: Select monitor mode using the Monitor function
button.
3. If necessary: Switch between parameter view and curve view
with the view button .
4. Following a shock delivery: Wait until the ECG has stabilized
again.