6 Operation
Result The ECG electrodes and ECG cable for 12-lead ECG recording and
assessment are connected. 12-lead ECG recording and assessment
is prepared.
6.11.2 Carrying out and delivering 12-lead ECG
recording and assessment
Requirement • The device is switched on (see “6.1 Switching on the device”,
page 118).
• A patient group is selected (see “6.4 Selecting patient group”,
page 120).
• 12-lead ECG recording and assessment is prepared (see
“6.11.1 Preparing 12-lead ECG recording and assessment”,
page 171).
1. Call up all ECG leads using the Lead function button:
• Check whether all ECG leads are being displayed in a stable
manner. If the ECG leads are not stable: Ensure the ECG
electrodes are connected to the patient and ECG cable.
• Check whether the sequence of ECG leads is plausible.
• Following a shock delivery: Wait until the ECG has
stabilized again.
2. If a 12-lead ECG has already been recorded in the current
session: Call up the last 12-lead ECG with the Last function
button.
When doing so, please note:
• You can only call up the last 12-lead ECG of the current
session.
• If no 12-lead ECG has yet been recorded in this session, the
Last function button will not be available.
3. To cancel 12-lead ECG function mode:
Cancel 12-lead ECG function mode using the Finish function
button.