WM 68401 04/2021
MEDUCORE Standard
2
EN 181
6 Operation
14. If desired (only with E-mail delivery 12-lead ECG option): Press
the Send function button.
15. Select the recipient's name and send the 12-lead ECG to a
predefined recipient by e-mail (see “12.5 Configuring e-mail
delivery 12-lead ECG (only with E-mail delivery 12-lead ECG
option)”, page 299).
Following successful delivery, the symbol appears in the
display. If delivery is unsuccessful, the symbol appears in
the display and an alarm is displayed (see “11.2 Alarm
messages”, page 271).
Delayed therapy due to severely delayed reception of the
12-lead ECG!
If the infrastructure of the monitoring station/hospital is not
configured by the operator to receive 12-lead ECGs, receipt of the
12-lead ECG may be severely delayed. This may lead to delayed
therapy and injure the patient.
Configure e-mail delivery 12-lead ECG correctly and check it
(see “12.5 Configuring e-mail delivery 12-lead ECG (only with
E-mail delivery 12-lead ECG option)”, page 299).
If transmission time is unexpectedly high: Contact the system
administrator of the e-mail recipient.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals