6 Operation
A patient group is selected (see “6.4 Selecting patient group”,
page 120).
A non-invasive blood pressure measurement (NIBP
measurement) is prepared (see “6.12.1 Preparing non-invasive
blood pressure measurement (NIBP measurement)”,
page 182).
1. Press the NIBP button for < 2 s.
The device switches to NIBP function mode.
2. If the patient group is incorrect: Select another patient group
(see “6.4 Selecting patient group”, page 120).
The NIBP module is configured accordingly in the device with
the selected patient group.
Risk of injury due to incorrectly selected patient group!
The device only delivers correct measured values if the appropriate
patient group is selected. An incorrect patient group may lead to
incorrect measurements and injure the patient.
Adapt the patient group to the patient.
If the patient group is incorrect: Change the patient group in
the application menu.