6 Operation
3. Adapt the initial NIBP cuff pressure to the patient using the
Initial function button and navigation knob.
When doing so, please note: Following successful NIBP
measurement, the initial NIBP cuff pressure adapts to the
patient (approximately 30 mmHg above the systolic measured
value of the previous NIBP measurement).
4. Press the Start function button.
Alternatively:
Press the NIBP button for > 2 s.
An NIBP measurement starts. After the end of the
measurement, the device displays systolic and diastolic arterial
pressure.
5. Evaluate the NIBP measurement result.
6. If implausible measured values are shown:
Check whether the NIBP cuff is correctly selected and
attached.
If necessary: Repeat the NIBP measurement.
7. If necessary: Cancel the NIBP measurement with the Stop
function button.
The device releases the pressure from the NIBP cuff.