7 Disassembling
7.1.3 Disconnecting the device from the 12 V
on-board power supply
Requirement The device is switched off.
1. When used in a protective transport bag: Open the rear of the
device compartment of the protective transport bag.
2. Disconnect the charging adapter from the power supply
connection.
3. Disconnect the charging adapter from the adapter cable for
12 V on-board power supply/circular connector.
4. Disconnect the adapter cable for 12 V on-board power supply/
circular connector from the 12 V on-board power supply.
Result The device is disconnected from the 12 V on-board power supply.