WM 68401 04/2021
MEDUCORE Standard
2
EN 241
8 Hygienic reprocessing
Power information
Accu-Pack
battery
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
Adapter cable
for 12 V on-
board power
supply/circular
connector
Power supply
unit and charger
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Not necessary Not permitted Not permitted Not permitted
Charging
adapter
Charging station
for battery
Part
Manual
cleaning
(only
necessary in
the event of
visible dirt)
Disinfection
by wiping
Disinfection
by
immersion
Mechanical
reprocessing
Sterilization