EasyManua.ls Logo

Weinmann MEDUCORE Standard2 - Page 242

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
242 EN MEDUCORE Standard
2
WM 68401 04/2021
8 Hygienic reprocessing
Defibrillation/cardioversion
MCS2-Connect
master cable
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
MCS2-Hardpads
paddles
Function test
resistor
MCS2-Softpads
defibrillation
electrodes for
children/adults
(disposable)
Not permitted, as disposables
Electrode gel
(disposable)
Part
Manual
cleaning
(only
necessary in
the event of
visible dirt)
Disinfection
by wiping
Disinfection
by
immersion
Mechanical
reprocessing
Sterilization

Table of Contents

Related product manuals