242 EN MEDUCORE Standard
2
WM 68401 04/2021
8 Hygienic reprocessing
Defibrillation/cardioversion
MCS2-Connect
master cable
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
MCS2-Hardpads
paddles
Function test
resistor
MCS2-Softpads
defibrillation
electrodes for
children/adults
(disposable)
Not permitted, as disposables
Electrode gel
(disposable)
Part
Manual
cleaning
(only
necessary in
the event of
visible dirt)
Disinfection
by wiping
Disinfection
by
immersion
Mechanical
reprocessing
Sterilization