8 Hygienic reprocessing
Pulse oximetry monitoring
MCS2-Adapt
pulse oximetry
sensor
connecting
cable
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
MCS2-SoftTip
pulse oximetry
sensor
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
MCS2-Wrap
pulse oximetry
sensor
MCS2-Earclip
pulse oximetry
sensor
Ear clip for
MCS2-Earclip
pulse oximetry
sensor
MCS2-Wrap
pulse oximetry
sensors
(disposable)
Not permitted, as disposables
Strap for
attaching
MCS2-Wrap
pulse oximetry
sensor
(disposable)
Part
Manual
cleaning
(only
necessary in
the event of
visible dirt)
Disinfection
by wiping
Disinfection
by
immersion
Mechanical
reprocessing
Sterilization