244 EN MEDUCORE Standard
2
WM 68401 04/2021
8 Hygienic reprocessing
6-lead ECG monitoring/12-lead ECG recording and assessment
MCS2-Line ECG
cables (all
versions)
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
ECG cable
separator
ECG electrodes
for adults and
children
(disposable)
Not permitted, as disposables
Non-invasive blood pressure measurement (NIBP monitoring)
NIBP connecting
tube
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dose:
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Not permitted Not permitted Not permitted
Adapter tube for
connecting NIBP
disposable cuffs
for newborns
NIBP cuffs
(reusable)
Follow the instructions for use of the NIBP cuffs
NIBP cuffs
(disposable)
Not permitted, as disposables
Part
Manual
cleaning
(only
necessary in
the event of
visible dirt)
Disinfection
by wiping
Disinfection
by
immersion
Mechanical
reprocessing
Sterilization