274 EN MEDUCORE Standard
2
WM 68401 04/2021
11 Alarms and faults
NIBP diastolic r
(Alarm deactivated after 10 s)
Measured diastolic blood pressure
is below the set lower alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits (see 10.3.1, p.
255).
NIBP measurement error
(Alarm deactivated after 10 s)
NIBP measurement cannot be
carried out
Make sure that the patient lies
still during the measurement.
Rule out the risk of patient
vibration.
Use a suitable size of NIBP
cuff.
Ensure that the NIBP cuff is
properly attached.
Ensure that there is no
clothing between the NIBP
cuff and the patient.
NIBP module defective
(Alarm active as long as the
condition prevails)
Internal NIBP module defective Have the device repaired.
NIBP outside measuring range
(Alarm deactivated after 10 s)
Blood pressure cannot be
measured and displayed as it is
outside the device’s measuring
range
Use an alternative blood pressure
measurement device with a larger
measuring range.
No NIBP signal
(Alarm deactivated after 10 s)
NIBP module unable to detect a
pulse wave signal
Make sure that the patient lies
still during the measurement.
Rule out the risk of patient
vibration.
Use a suitable size of NIBP
cuff.
Ensure that the NIBP cuff is
properly attached.
Ensure that there is no
clothing between the NIBP
cuff and the patient.
NIBP safety shutdown
(Alarm deactivated after 10 s)
Pressure in the NIBP cuff is too
high for too long
Remove the NIBP cuff from the
patient. Have the device repaired.
Alarm Cause Remedy

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals