EasyManuals Logo

Weinmann MEDUCORE Standard2 User Manual

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #273 background imageLoading...
Page #273 background image
WM 68401 04/2021
MEDUCORE Standard
2
EN 273
11 Alarms and faults
ECG module defective Internal ECG module defective Have the device repaired.
E-mail delivery error
(Alarm deactivated after 60 s)
Server not accessible
Check connection to network (see
12.3, p. 292).
Connection between device and
network interrupted
Check preset e-mail recipient (see
12.5, p. 299).
Check connection to network (see
12.3, p. 292). If necessary: Select a
different network.
Check date/time (see 10.3.7, p.
264).
Excessive NIBP cuff pressure
(Alarm deactivated after 10 s)
Pressure in the NIBP cuff increased
to such a degree as a result of
external influences that the safety
valve opened
Check which influences resulted in
the pressure increase and
eliminate them before the next
measurement.
Heart rate q
Measured heart rate is above the
set upper alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits (see 10.3.1, p.
255).
Heart rate > 250/min
Measured heart rate is above 250/
min (device shows --- as heart
rate)
If medically indicated, carry out
suitable treatment.
Heart rate r
Measured heart rate is below the
set lower alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits (see 10.3.1, p.
255).
Heart rate < 30/min
Measured heart rate is below 30/
min (device shows --- as heart
rate)
If medically indicated, carry out
suitable treatment.
Insert printer paper
Printer paper completely used up
Change printer paper (see 4.7.3, p.
98).
No printer paper inserted
Change printer paper (see 4.7.3, p.
98).
Memory error Internal memory module defective Have the device repaired.
NIBP diastolic q
(Alarm deactivated after 10 s)
Measured diastolic blood pressure
is above the set upper alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits (see 10.3.1, p.
255).
Alarm Cause Remedy

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals