282 EN MEDUCORE Standard
2
WM 68401 04/2021
11 Alarms and faults
11.3.4 6-lead ECG monitoring/12-lead ECG recording
and assessment
Fault Cause Remedy
The Plug in ECG connector
alarm occurs although the ECG
cable is connected to the device
Device not detecting connected
ECG cable
Check plug connection.
Change ECG cable.
Have the device repaired.
The ECG cable is not connected to
the device but the Plug in ECG
connector alarm does not occur
Device incorrectly detecting ECG
cable as connected
Have the device repaired.
The alarm only occurs if an ECG
has already been successfully
derived since switching on
-
The Check seating of ECG
electrodes alarm occurs
although the ECG electrodes are
attached to the patient
ECG electrodes not attached
correctly to the patient
Attach ECG electrodes to the
patient correctly (see 6.10.1, p.
164).
Change ECG cable.
Have the device repaired.
ECG cable not attached/not
attached correctly to the patient,
but the Check seating of ECG
electrodes alarm does not occur
Device incorrectly detecting ECG
cable as attached to the patient
Have the device repaired.
The alarm only occurs if an ECG
has already been successfully
derived since switching on
-

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals