11 Alarms and faults
11.3.3 Defibrillation/cardioversion
Battery not charging although it is
not full
Battery temperature < 0 °C or
> 45 °C
Charge battery within permitted
temperature range (see 18.5, p.
371).
Battery defective Replace battery.
Status LEDs are not flashing and
battery is not charging
Battery not inserted correctly in
charging station
Check battery.
Insert battery correctly.
Charging station not connected to
the power supply
Connect the charging station to
line power (see 4.5.3, p. 90).
Connect the charging station to a
12 V on-board power supply (see
4.5.4, p. 90).
Charging station defective Have charging station repaired.
Battery capacity at > 90 %
Press the status button on the
battery to check battery status.
Fault Cause Remedy
Fault Cause Remedy
The Check defibrillation
electrodes alarm occurs
although the defibrillation
electrodes are attached to both the
patient and the device via the
master cable
Defibrillation electrodes not
attached correctly to the patient
Attach defibrillation electrodes to
the patient correctly (see 6.5.1, p.
122).
Defibrillation electrodes defective Replace defibrillation electrodes.
Pad connector not connected
correctly to the master cable
Connect Pad connector of the
defibrillation electrodes to the
master cable correctly.
Master cable not connected
correctly to the device
Connect the master cable to the
device correctly.
Master cable defective Replace the master cable.
Defibrillation module defective Have the device repaired.