Chapter 5 EDITING AUTOMATIC SEARCH RESULTS
5-31
Changing the Splice Loss Threshold
For details on splice loss threshold, refer to page 3-42.
1.
Locate the cursor to “SPLICE LOSS” by using the rotary knob or the arrow key.
2. Press [ENTER].
A selection window will appear.
3. Change to the desired splice loss threshold.
For method of changing the splice loss threshold, refer to page 3-42.
4.
Press [F1] (Re-SEARCH START).
Auto search will start with the newly set search conditions.
When changing the condition for two or more items, it is recommended to change
the condition for each item first and then press [F1].
Refer to the notes given on changing the search conditions (page 5-38).
Refer
Refer
TIP
Refer