Solution Volume Display
AEROSET
®
Customer Training Guide 3-21
94858-103 — July 2003
ICT™ Reference Solution Loading
Procedure
CAUTION: The ICT Reference Solution should not be
changed when the System status is RUN, PAUSE, or
SAMPLING END because the System is using the
solution at these times and sample results could be
affected.
NOTE: If the “88 - Inventory Empty, ICT Reference
Solution” Error Log Message is generated during a run,
the System automatically switches to PAUSE status. No
additional samples are aspirated but samples already
dispensed continue to process. The Operator can:
Allow these samples to complete.
•Select <STOP> to change the System status to
READY.
•Select <UNPAUSE> to continue sample aspiration
of photometric assays only.
ICT results completed after the Error Log Message is
generated should be rerun.
NOTE: Refer to Error Log Messages—Supplemental
Information in Section 10, Troubleshooting and
Diagnostics of the AEROSET System Operations
Manual for details on initial recovery from Error Log
Message 88.
1. Ensure the System status is READY prior to loading the ICT
Reference Solution.
2. Open the front-right door of the Analyzer.
3. Remove the used ICT Reference Solution bottle from the tray.
4. Remove the reference pump tubing and associated cap from the
used bottle.
5. Prepare the new ICT reference solution by removing the shipping
cap from the bottle.
!