MODULE 3: DAILY START UP AND CHECKS
3-22 AEROSET
®
Customer Training Guide
94858-103 — July 2003
6. Insert the ICT reference pump tubing in the new bottle and seat the
cap on the bottle opening. Firmly press down on the cap to secure it
on the opening.
NOTE: Ensure the cap covers the ICT reference solution
bottle threads on the neck of the bottle, and the tubing is
inserted to reach the bottom of the bottle.
CAUTION: Failure to comply could cause aspiration of
air and erroneous results.
7. Place the bottle securely on the tray. Verify the following
conditions exist:
Area is clean and free of debris
Bottles are seated properly
Upward movement of the wash solution tray is not
restricted
8. Close the front-right door of the Analyzer.
9. Select <START UP> in the Action Area of the Main Display. The
START UP OPTIONS screen displays.
10. Select the checkbox (5
) next to the Wash ICT with IRef and
Drain & Fill IRef Cup options. Ensure all other options are
deselected ().
11. Select <Start> in the upper-right corner of the START UP
OPTIONS screen to begin the procedure.
12. When START UP is complete, perform an ICT calibration and run
all levels of QC.
!