MODULE 3: DAILY START UP AND CHECKS
3-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
2. Check each of the syringes in the pump. Visually inspect the syringe
interior, below the plunger tip, for evidence of leakage.
Replace the syringe if leaking is observed. For replacement of a
1 mL syringe, refer to Component Replacement in Section 9,
Service and Maintenance of the AEROSET System Opera-
tions Manual.
3. Open the ICT™ Pump access door to access the ICT Aspiration
Pump and ICT Reference Solution Pump 1 mL syringes.
Figure 3.7: ICT Aspiration Pump and ICT Reference
Solution Pump
125634
Check for
Leakage
ICT Aspiration Pump
ICT Reference
Solution Pump