Daily Checks
AEROSET
®
Customer Training Guide 3-27
94858-103 — July 2003
Check Sample and Reagent Syringes and Solenoid Valves
1. Open the front-middle door of the Analyzer.
2. Inspect the Sample and Reagent Syringes for leaks from the syringe
body.
3. Inspect the three (3) connections of the Sample and Reagent
Syringes for leaks as shown in the following illustration.
1 = Sample A Syringe Drive
2 = Reagent 1A Syringe Drive
3 = Reagent 2A Syringe Drive
4 = Sample B Syringe Drive
5 = Reagent 1B Syringe Drive
6 = Reagent 2B Syringe Drive
125635
3
2
1