MODULE 3: DAILY START UP AND CHECKS
3-26 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Check Wash Solution Trays
1. Open the front-right door of the Analyzer to access the Wash
Solution weight platforms.
2. Verify that no liquid has accumulated in the tray under the Wash
Solution bottles.
3. Check that the bottles have not overfilled, indicating that a solenoid
valve is defective. Contact your Abbott Representative if there is an
indication of overfill.
4. Close the front-right door.
CAUTION: Chemical Hazard. Refer to Appendix, Haz-
ards of this manual. Also refer to product-specific
Material Safety Data Sheets (MSDS).
!